S E P T E M B E R 1, 1936
PHYSICAL REVIEW
VOLUM E 50
The Gyromagnetic Properties of the Hydrogens
J. M. B. KELLOGG, I . I . R A B I * AND J. R. ZACHARIAS, Columbia University
(Received June 25, 1936)
The magnitudes of the nuclear moments of the proton and the deuteron are remeasured by the method of atomic beams. The new experimental arrangement is such that the evaluation of the results does not require any information with regard to the velocity distribution of the atoms in the beam. Detection is made objective by the use of a Stern-Pirani detector. The signs of the magnetic moments of the proton and deuteron are determined by the method of nonadiabatic transitions in a weak magnetic field. The results of the experiment are: MP = 2 . 8 5 ±0.15; ,uo = 0.85 ±0.03 nuclear magnetons. Both moments are positive in sign.
THE discovery by Stern1 and his collaborators that the magnetic moment of the proton is very much larger than the value predicted by the Dirac theory suggests a number of important problems. These include the development of methods for a more precise determination of the
nuclear moment depends, which makes the result in principle independent of any assumption as to the velocity distribution of the atoms in the beam. This eliminates an important source of possible error in measurements of this kind.
magnitude of the moment and the question of the sign. Because of its much smaller moment the same questions arise in more aggravated form for the deuteron.
The question of the sign of the nuclear moments is particularly interesting since the proton is so anomalous that considerations from
GENERAL CONSIDERATIONS
A hydrogen or deuterium atom in the normal 2S$ state may have in a magnetic field one of several values of atomic4 magnetic moment /.5 For hydrogen these moments are: (in units of Ho, the Bohr magneton)
the Dirac theory do not suffice for its prediction, while the deuteron is the only nucleus of its type
=bfi=±l,
(1)
(even atomic weight, odd nuclear charge) for which the moment has been measured,2 and and for deuterium the moments are:
there is therefore no example from which to
±/i=±l,
draw an analogy. This question arises only when the moment is measured by the deflection of molecular or atomic beams,3 since it is peculiar to these methods that the sign of the nuclear moments does not affect the deflection pattern.
In the experiments to be described we have developed and applied to hydrogen and deuterium a method of determining the sign of the nuclear moments which is based on the use of
±/2=±(x+i)/(l+fx+x2)*,
(2)
=b/3=±(x-i)/(l-fx+x2)i
The moments are plotted against x in Figs. 1A and B. The parameter x is defined by x = 2fjLoH/hcAv, where II is the value of the magnetic field. The quantity Av is the separation in cm-1 of the two hfs components in zero field and is related to the nuclear magnetic moment HN by
nonadiabatic transitions between states of space
A^ = [(2i + l ) A ] ( 8 / 3 ^ ) M ^ o ^ 2 ( 0 ) , (3)
quantization of the atom in weak magnetic fields. We have also developed a method of measuring the hfs separation of the normal state of the atom, on which the evaluation of the
where i is the nuclear spin and ^(0) is the value
of the Schrodinger eigenfunction at the nucleus
which can be calculated exactly. Both from band spectra determinations6 and, as will be apparent
* Ernest Kempton Adams Fellow, Columbia University, 1935-36.
later, from the experiments here described, it is
Arisen and Stern, Zeits. f. Physik 85, 4 (1933); Ester- known that the nuclear spin of hydrogen is
mann and Stern, Zeits. f. Physik 85, 17 (1933). 2 Estermann and Stern, Phys. Rev. 45, 761 (1934); Rabi,
4 Atomic magnetic moment is not to be confused with
Kellogg and Zacharias, Phys. Rev. 45, 769 (1934).
nuclear magnetic moment.
3 Rabi, Kellogg and Zacharias, Phys. Rev. 46, 157, 163 5Breit and Rabi, Phys. Rev. 38, 2082 (1931).
(1934).
6 Murphy and Johnson, Phys. Rev. 45, 761 (1934).
472
G Y R O M A G N E T I C P R O P E R T I E S OF H Y D R O G E N
NUCLEAR MOMENT
DEUTERIUM
POSITIVE NEGATIVE
f,
473
AA' x*
.5
/. 0
xx -
*
H
x2-
>** F-yz v*\--*\ F^fe ^=»>/2 - \
FIGS. 1A and IB. Diagram to show quantum numbers associated with the magnetic states assuming the nuclear spin positive or negative.
\ and the nuclear spin of deuterium is 1. For hydrogen, with nuclear moment fxp expressed in units of /*o/1838, the above equation reduces to Av = 0.0169JJ,P, and for deuterium with nuclear magnetic moment MD expressed in the same units, to Av' = 0.0127fin- Thus if an experimental evaluation of one of those atomic moments which depends upon x can be made while the atom is in a known field Hf the magnitude of the nuclear moment can be determined.
The results of a deflection experiment may be made to give both of these quantities. Consider a narrow beam of atoms emerging from the collimating slit of Fig. 3 and passing down the apparatus toward the slit of the Stern-Pirani detector. These atoms pass first through a region A of weak inhomogeneous magnetic field and then through a second field B similar to the first but stronger and arranged to give deflections in the opposite direction to those produced by the field A. This latter condition may easily be satisfied since atoms of positive moment are always deflected into regions of stronger field. The deflection of an atom in the direction y perpendicular to the beam is given by
s=(l/4E)n(dH/dy)L,
(4)
L = l i 2 + 2 1 i l 2 , with E the kinetic energy of the atom, li the distance the atom moves in the
field, and 12 the distance the atom traverses from the end of the field to the detector. A suitable choice of H and dH/dy in the fields A and B will give equality of the deflections produced in these regions, and one can write independently of the value of E and hence of the velocity distribution
f(dH/dyYLf=f"(dH/dy)"L"}
(5)
where the primes and double primes refer to the values of the various quantities in the fields A and B, respectively. If the values of H and dH/dy in the two fields are known, this equation gives immediately the value of the nuclear moment. In our apparatus the quantities H and dH/dy are proportional to and can be immediately calculated from the currents used to produce the magnetic fields. With / ' and I" in amperes approximate values of H and dH/dy for the two fields may be found from H' = 1.34J', H" = 1.16 J" and (dH/dyy = llAI', (dH/dy)" = 5.951".
A measurement of the atomic moments according to the theory outlined above is sufficient to determine the magnitude of the nuclear magnetic moments, but because of the symmetry of the deflection patterns cannot yield information as to whether this moment is positive or negative, i.e., whether the nuclear magnetic moment is oriented parallel or antiparallel to the
474
KELLOGG, RABI AND ZACHARIAS
TABLE I. Transition probabilities for F—l.
TABLE II. Transition probabilities for F=3/2. c = cos [a/2) ,s = sin ( a / 2 ) .
m
-1
-1
c4
0
2s2c2
+ 1
s4
0 2c2s2 l-4^2 2c2s2
+ 1
s4 2s2c2
c*
m
-3/2 -1/2 + 1/2 + 3/2
-3/2
c6 3c*s2 3c2s*
-1/2 3cAs2 c2(c2-2s2)2 s2(2c2-s2)2 3c2s*
+ 1/2 3c2s* . s2{2c2-s2)2 c2{c2-2s2)2 3c*s2
+3/2 3cAs2
cG
nuclear angular momentum vector. This question
has been discussed by Rabi in a paper "On the Process of Space Quantization,"7 and he there
points out that certain nonadiabatic processes
may be used to discover the sign of the nuclear
moment.
velocity. We cannot, therefore, begin by setting the field at some particular value and expect to find nonadiabatic transitions. The procedure which we followed was to find experimentally a value of the current in T for which the transitions
Consider an atom of the beam moving with occurred.
constant velocity through a magnetic field The effective value of a can be calculated from
varying in strength and direction along its path. our knowledge of the fraction of the atoms which
The atom is equivalent for these questions to an atom at rest and situated in a field varying in time in the same manner. If the angular velocity of rotation of the field, T, is small compared
do not make the transitions and the value of m for the state which has been selected. Since the atoms in the other states pass through the same field and have the same velocity distribution,
with the Larmor frequency co = 2wnogH/h the they have the same effective a. We are thus at atom will remain space quantized with respect to liberty to use a field for which the dynamical the field with the same component m of its total problem has not been solved. The type of field angular momentum F (adiabatic transforma- which we adopted for reasons of convenience is bility); if the angular velocity is of the same shown in Fig. 2A. This field is neither the order of magnitude as the Larmor frequency Giittinger9 nor Majorana type although nearer there will be nonadiabatic transitions to states the former. Data obtained in this manner cannot mf not necessarily equal to m but with the same be used to exhibit the quantitative side of the F. The probabilities of such transitions have been dynamical theory of these transitions. Our calculated by Majorana8 and are tabulated in purpose, however, is to recognize a qualitative
Tables I and II for hydrogen and deuterium. The parameter a which occurs in these expressions is obtained from the dynamical theory of the process. Physically a is the angle between the original direction in which the atom is space quantized (direction of the field H) with quantum number m, and the direction after the process with respect to which the atom is quantized with the same quantum number m. Majorana has shown that this process depends only on the Lande g factor and the nature of the process, and
difference in the behavior of the atoms in the two F states under identical conditions.
Reference to Figs. 1A and IB shows that the assignment of the quantum numbers m and F to atoms in the several magnetic states depends on the sign of the moment of the nucleus. It is this dependence, together with the nonadiabatic transitions made use of in a method to be described, which enables a decision to be formed as to the sign of the nuclear moments.
is independent of m. Since the two F states have the same g value, the value of a is the same for
APPARATUS
both for any dynamical process whatsoever.
The arrangement of the pumps and the slit
Since the form of the field T is not known exactly it is not possible to calculate a as a function of the field and of the velocity. There will also be a different value of a for every atomic
system (Fig. 3) is almost the same as that described by R. K. Z.3 In the present apparatus
the beam height is 2 mm, and the widths of the
various slits are: source slOt i.03 mm; fore slit,
0.05 mm; collimating slit, 0.02 mm; selector
7 Rabi, Phys. Rev. 49, 324 (1936). 8 Majorana, Nuovo Cim. 9, 43 (1932).
9 Giittinger, Zeits. f. Physik 73, 169 (1931).
G Y R O M A G N E T I C P R O P E R T I E S OF H Y D R O G E N
475
FIG. 2. Diagrams of field near wires T. (A) Nonadiabatic field; (B) Adiabatic field.
slit, 0.5 mm; and detector slit, 0.01 mm. Approximate distances measured from the source slit to the points indicated are: fore slit, 1.0 cm; collimating slit, 17 cm; end of A field, 33 cm; beginning of B field, 38 cm; end of B field, 48 cm; and detector, 50 cm.
The discharge tube: The hydrogen atoms are prepared in a long Wood discharge tube of a design (see Fig. 4) which permits the source slit to be in close proximity to the discharge. The slit is formed by ground edges of thin cover glass held on to the Pyrex tube with picein wax. Water cooling prevents melting of the wax and also seems to keep the temperature of the gas fairly low. With gas pressures of the order of 1 mm the concentration of hydrogen atoms in the beam is between 0.7 and 0.9. In the previous experiment3 this ratio was only 0.1 to 0.2. The concentration of atoms is taken from measurements of full beam intensity and of the intensity with enough deflecting field on to cast out all atoms.
The ground joint in the threaded collar of Fig. 4 is bored with its axis eccentric and at 5° to the axis of the screw threads. This permits alignment of the source slit parallel to the collimator slit. Motion of the source slit in the horizontal direction is accomplished by moving the brass base plate or by turning the source tube. The canal of the source slit can thus be made to point toward the collimator slit. With the aid of these adjustments it is possible to replace the discharge tube after removing it from the apparatus for cleaning and putting on new slit jaws. This procedure was necessary because Apiezon oil from the diffusion pumps sometimes clogged the source slit.
The magnetic fields: The A and B fields are produced in the manner previously described3 by current flowing in two horizontal straight tubes parallel to the beam. Great care was exercised both in the construction of the ''fields" and in their proper placement with respect to the beam. The B field differs from the A in that it is shorter, capable of carrying more current, and gives a smaller ratio of gradient to field.
Although rheostats, switches, etc., are not worth describing, the leads for the field supply are noteworthy in one regard: in the reduction of stray field. Parallel bus bars of copper placed close together suffice for distances of 20 cm or more from the beam. The insert of Fig. 3 represents a cross section of the leads to the current tubes of the second field in the region near the beam. The cross hatched copper tube is slotted and is soldered to the tube which carries
TO BATTERY u AND WATER SUP°LY
FIG. 3. Diagram of apparatus as used for determination of the sign of the moment for hydrogen and deuterium. The permalloy shield and the wires T were removed from the apparatus for the measurement of the magnitudes of the moments.
476
KELLOGG, RABI AND ZACHARIAS
WATER LEADS TO BOTTOM OF WATER JACKET
FIG. 4. Section of source chamber to show the construction of discharge tube and of the eccentric brass adjusting collar.
the current in; the central copper tube carries the current out. This arrangement permits simple water connections, simple vacuum seal, good conductivity and small stray field.
A parallel bank of 12 lead storage cells with a capacity of about 1500 amp.-hrs. supplies the first field. For the B field, which requires about 800 amp. continuously, a bank of 6000 amp.-hrs. is maintained. These batteries are charged by a 12-volt, 1200-amp. generator.
The transition field: This field is produced by current flowing in-two vertical wires on one side of the beam. It is evident from Fig. 3 that there are four symmetrically disposed copper wires in a region magnetically shielded by permalloy. A system of switches, rheostats, batteries and meters makes it possible to use pairs of wires on
either side of the beam as the transition field wires, or to produce an adiabatic field by using one wire on each side of the beam. The forms of these two types of field with respect to the beam direction are shown in Figs. 2A and 2B. Magnetization of the permalloy shield is avoided by running the current in the wires in opposite directions through one hole in the top of the shield.
The detector: The Stern-Pirani detector is of the type designed by Zabel.10 The detecting gauge and the compensating gauge are enclosed in the same brass block and equipped with slits 2 mm high, 0.01 mm wide, and 4 mm deep. The motions necessary for its proper alignment with the beam are simple screw devices of obvious design. In the early part of the experiment considerable difficulty was encountered in reading the gauge galvanometer while the hydrogen discharge tube was in operation. Unsteadiness of the gauge seemed to be associated with fluctuating pressure in the detecting chamber and with subsidiary electrical effects of the discharge. Therefore the electrical circuit from the gauge wires to the galvanometer was shielded with sheet copper and the backing pressure for the high speed pumps was improved so as to maintain an ion gauge pressure of 10~~7 mm in the detecting chamber. To avoid electrical and mechanical disturbances by mechanical pumps a three-stage Leybold mercury diffusion pump was inserted in the fore line and the hydrogen gas from the source slit was pumped into an evacuated 30 gal. tank.
METHOD: MAGNITUDE OF THE MOMENTS
The detector slit is set to receive the atoms of the undeflected beam and is left fixed in this position throughout the course of all experiments. The simplest experiment to perform is the determination of the nuclear spin of hydrogen. The current in the B field is turned on and set at such a value as to give an x in the neighborhood of 1.5. With no current in the A field the beam is split in the B field and few atoms enter the gauge. If the A field current, I', is now increased, all atoms in the A field experience a deflecting force proportional to their atomic moments and to
' Zabel, Phys. Rev. 46, 411 (1934).
G Y R O M A G N E T I C P R O P E R T I E S OF H Y D R O G E N
477
the current V. This force is opposite in direction to the force that will act on them when they pass through the B field. At the same time atoms having moments other than ± / i are receiving larger deflections in the A field on two counts; first because the larger V increases H' and hence x and /2, and second because the force per unit moment is becoming larger due to the increase in (dH/dy)'. These atoms may thus in turn be focused into the detector. Focusing is of course recognized by a maximum of intensity at the detector.
A curve showing the two intensity maxima so obtained for hydrogen is given in Fig. 5. Since four atomic magnetic levels are required to give the two peaks of Fig. 5, and since the number of these levels is given by ( 2 i + l ) ( 2 j + l ) , a nuclear spin of \ is found for hydrogen.
While the existence of two and only two peaks is sufficient to establish the proton spin, a determination of the position (in current) of the peaks and a more detailed analysis of the operation of the apparatus is necessary before a value can be assigned to the magnitude of the magnetic moment of the proton.
First, the values of H and dH/dy are neither exactly constant over the height of the beam, nor because of end effects, along the beam. The terms f(dH/dy)L in Eq. (5) are therefore replaced by a numerical integration over the full path of the beam using values of H and dH/dy averaged over the height of the beam.
Second, the values of H and dH/dy are calcu-
lated from measured geometry of the apparatus, e.g., the distance between the wires, their diameters, the distance from the wires to the beam, and the lengths of the beam path both in and out of the fields. It is fortunately possible to use the values of the currents required to deflect and focus atoms of constant moment to evaluate directly some of the constants in Eq. (5). If we let dH/dy = GI, then Eq. (5) can be written
/ ' / ' / / "J" = G"L"/G'L'=R.
(6)
This ratio R can be evaluated in two ways. First it can be calculated directly from the geometry of the apparatus. Second, it can be determined in terms of the currents required to focus hydrogen atoms having / / = / / / = MO- A comparison of the values of R obtained in these two ways serves as a check on the geometry used in the calculation of H. The second determination is considered to be more accurate than the first and is used in the evaluation of the ratio f/'f" f ° r
/V/".
Third, consider two atoms emerging from the collimating slit, one with moment + / and the other with moment —/. The atom with moment + / is deflected in the A field into regions of stronger field and gradient while that with moment —/ is deflected into regions of weaker field and gradient. The result is to give the atom with moment + / a larger deflection than the atom with moment —/. The B field which is on the opposite side of the beam from the A field
HYDROGEN
10
50
100
150
200
CURRENT IN FIRST FIELD IN AMPERES
FIG. 5. Intensity at center of beam against field current for hydrogen to show nuclear spin to be f. The first peak is for i / i ; the second for db/2.
478
KELLOGG, RABI AND ZACHARIAS
1(^
B-FIELD CURRENT
823 AMP.
llSO 11 60 170
180
190
200
CURRENT IN FIRST FIELD IN AMPERES
FIG. 6. Intensity of beam against field with selector slit interposed to pass only atoms with negative moments, for hydrogen.
gives the amount by which the intensity at one current exceeds that at the other. By interrelating two points on the curve so directly one largely eliminates effects of secular changes in the intensity of t h e beam a n d changes in t h e ratio of t h e number of atoms to molecules. All current-intensity curves were taken in this manner.
METHOD: SIGN OF THE MOMENTS
Hydrogen
magnifies this effect and + / is underfocused whereas —/ is overfocused. T h u s instead of a sharp peak a broad maximum is formed by the superposition of t h e two partially focused beams.
This difficulty is overcome in the following manner. A selector slit is introduced where the beam is split between t h e two fields. By means of this slit it is possible t o c u t o u t of t h e beam either those atoms with positive or those with negative moments. This slit serves the additional purpose of blocking t h e undissociated molecules of t h e direct beam from t h e detector. Suppose the slit is set to pass only those atoms having positive moment. An intensity-current plot is made for, say, atoms of moment + / (see Fig. 6). The slit is then moved to pass only those atoms having negative moment and an intensitycurrent plot made for atoms of moment —/. If the positive moment peak occurs at current 1+', and t h e negative moment peak a t /_/, then (I+' + I-')/2 is very closely t h e current a t which the peak would occur in an ideal apparatus, i.e., an apparatus in which the deflecting force on an a t o m is n o t a function of t h e deflection.
Fourth, a method of using t h e detector is required which will give as accurately as possible the current at which the maximum intensity occurs. A ''differential'' method of measuring t h e intensity is adopted. With a given B field current, the current in t h e A field is set in the neighborhood of t h e peak. T h e reading of the detector galvanometer is noted and simultaneously a switch is thrown to vary the A field current b y a few percent. After thirty seconds the galvanometer reading is again noted and the switch thrown t o give t h e original A field current. This procedure is continued for six minutes and the readings averaged t o eliminate t h e effect of drift. T h e difference in t h e two sets of readings then
T h e transition field T is fixed in place between fields A a n d B. T h e selector slit is set t o pass atoms of positive moment a n d t h e A a n d B field currents adjusted to focus t h e component + / 2 . According t o Fig. 1A this component is
either J F = 0 , ra = 0, or F=l, tn = Q, depending on
the sign of t h e nuclear moment. If t h e state is F=l, m = 0 (nuclear moment negative), then Table I shows t h a t for a certain value of a, t h a t is, a certain value of t h e current in the wires of the field T, there will be nonadiabatic transitions to states F=l, m = ± l . If the transition is to m— + 1 , t h e a t o m enters t h e B field region with the sign of its moment changed a n d is no longer focused. There is therefore a decrease in intensity a t t h e detector. If t h e transition is to m = — 1 no decrease in intensity will be observed since t h e value of x in t h e second field is so large that all atoms experience approximately the same deflecting force. In a n y discussion of these experiments, a n d particularly in t h e case of deuterium, this equality of atomic moments in the B field must be kept in mind. In the vernacular of the laboratory "only those transitions t h a t cross t h e diagram (Fig. 1) are counted as 'flop' (nonadiabatic transition)." However, if t h e state is F—0, m = 0 (positive nuclear m o m e n t ) , no variation in intensity will be observed since changes in F are forbidden. A similar argument may easily be carried out for —/2. T h e results are summed up in Table III where the expected change in intensity is given for the various possible combinations of nuclear a n d atomic moment.
While it would seem sufficient to perform an experiment on, say, -f/2 and to determine whether this state does or does not make transitions, it is experimentally necessary to show that the result obtained is not due to
GYROMAGNETIC P R O P E R T I E S OF HYDROGEN
479
extraneous effects of the apparatus. T h u s , if it is found t h a t for a particular value of current in the T field wires there are no transitions made from the component +J2, it is then necessary to show t h a t for the same value of the T field there are transitions made from the component — ft. Furthermore a similar experiment on the components ± / i should show t h a t both of these components do make transitions. Lastly, to be sure that the asymmetry is not inherent in the position of the beam with respect to the field wires, this whole series of experiments must be performed with the transition wires located on the other side of the beam.
Deuterium
DEUTERIUM
B FIELD 349 AMP.
o°_o1oo oo
100
1
o
o o
1 oo
? o
60
1
?
T
O
O
120
40
60
80
CURRENT IN FIRST FIELD IN AMPERES
FIG. 7. Intensity of beam against field with selector slit set to pass only atoms with negative moment.
T h e same a p p a r a t u s is used as for the sign inside the permalloy shield of the transition
determination for hydrogen. T h e selector slit field, it is not certain t h a t mere absence of
is set to pass atoms of negative moment, and the current in these wires gives an adiabatic field in
A and B field currents are adjusted so t h a t this region. T h e very existence of the ferro-
only states —/2 and —/3 enter the detector. As is magnetic material of the shield in this space
obvious from Fig. 7, when —/3 is focused, the gives rise to the possibility of residual fields
beam is contaminated b y —f%. T h e procedure which might accidentally fulfill the requirements
followed is to focus partially both —/3 and —f^. for a nonadiabatic field. However, with current
Suppose the nuclear m o m e n t positive. T h e n flowing in opposite directions in wires 1 and 3
transitions are possible to all states with F=3/2, the field will be as shown in Fig. 2B and the
and in particular to the state F=3/2, m— —3/2, shield and the region inside t h e shield will
i.e., to + / i . Detailed consideration of Table II certainly be magnetized in such a manner t h a t
shows t h a t for an equal mixture of atoms in there will be no rapid change of direction of the
states —ji and —fs this transition probability is a maximum for sin2 (a/2) = ^. Therefore if the
field along the p a t h of the beam. All observations here reported were made by repeatedly switching
nuclear m o m e n t is positive, there will be a value from configuration B to configuration A of Fig. 2.
of current in the T field wires which will produce
If this procedure is not followed, the results
a decrease in the intensity of the beam composed are essentially the same. However, there were
of a t o m s of moments —f2 and —/3.
slight increases of intensity on switching from no
On t h e other h a n d if the nuclear m o m e n t is current to configuration B which would indicate
negative, reference to Fig. I B shows t h a t the the presence of this accidental nonadiabatic field.
states —/2 and —/3 both have F—\ and therefore
transitions are possible only between these two
TABLE III. The qualitative expected change of focused beam intensity for nuclear moment assumed positive and
states. Remembering t h a t the atomic m o m e n t s negative; and the observed change in intensity.
in the second field are all practically equal to one
Bohr magneton, it follows that the total number of atoms focused into the detector is not changed thereby.
For the atoms with positive moments + / 2 and
COMPONENT FOCUSED
EXPECTED CHANGE OF INTENSITY
Positive Nuclear Moment
Negative Nuclear Moment
OBSERVED
CHANGE OF INTENSITY
+ / 3 a similar argument leads to opposite conclusions. T h e results of this discussion are summed up in Table III.
Hydrogen
Decrease Decrease Decrease None Decrease None
Decrease None Decrease
While current flowing in opposite directions in either wires 1 and 2 or 3 and 4 gives a nonadiabatic field over the region of the beam
Decrease
Deuterium +/* & +/> None
-ft & - / a Decrease
Decrease Decrease
None
Decrease None
Decrease
480
KELLOGG, RABI A N D ZACHARIAS
20
0
O 0
10
o
o
1 0
x—x X - * _ _ X
|
HYDROGEN
CIRCLES - \ CROSSES+4
TOTAL INTENSITY OF - i , 30 D!V
| OF +\ S5 0IY
20
10
o _o
o o
°
Lx.^
o ' ? °
*—i—*
1
DEUTERIUM
'
CIRCLES - ^ , - 5 j
CROSSES + $2 ,+S3
I0IAL INILNSII Y — o r - ^ , - * , 42 DIV 1 OF +% + i 48 DIV
gL 11 9—}L6 '
CURRENT I N WIRES 3AND 4 IN SERIES. AMPERES
FIG. 8. Nonadiabatic transitions.
RESULTS
T h e sign of the nuclear m o m e n t s : Typical experimental results are given in Fig. 8. They show the change in intensity on the introduction of the nonadiabatic field. T h e striking differences between the behavior of t h e states with positive atomic moments and those of negative atomic moments is very apparent. This difference is not due to a geometric asymmetry because the same behavior is obtained with either pair of wires 1 and 2 or 3 and 4. Consultation of Table III shows that the experiments establish the signs of t h e nuclear moments of hydrogen and deuterium to be positive.
T h e magnitudes of the nuclear m o m e n t s : Figs. 6 and 7 show typical experimental curves from which the intensity-current maxima were obtained. Table IV gives the results of a series of measurements of t h e q u a n t i t y R of E q . (6) for different values of the B field current and the corresponding values of t h e A field current for the low current peak. They fall within one percent of constancy and seem to show no trend with current. This table shows the precision of the determination of t h e maximum of intensity with current. T h e value of R calculated from the geometrical measurement of field wire distances etc. is 11.05 as compared with 11.09 which is the average of t h e values of Table IV. This check permits us to place confidence in t h e values of t h e moment as calculated from the currents in the A field necessary to focus the atomic moments =b/2.
T a b l e V gives t h e results of a series of measurements of these quantities and the proton moments calculated from them. Table VI gives the
TABLE IV. Hydrogen. 1+ is the current to focus +/i>
and / _ ' is the current to focus —f\. IA —
{I+JTI-,)/2.
Currents in amperes.
B Field Current /"
400 500 627 648 648
A Field Current
V
/_'
IA
32.1 40.1 36.1 39.5 50.3 44.9 50.8 62.5 • 56.7 52.0 65.5 58.7 50.9 65.5 58.2
R = I"/IA
11.08 11.13 11.06 11.03 11.13
TABLE V. Hydrogen. 1+ is the current to focus 4-/2, and /_' is the current to focus —f-2. IA— (/+' + ^ - 0 / 2 . Currents in amperes.
B Field Current
A Field Current
/"
/+'
/_'
IA
MP
601
121.7 142.7 132.2
2.79
732.5
140.0 166.0 153.0
2.87
732.5
141.5 164.5 153.0
2.87
773
145.7 170.0 157.7
2.81.
783
148.0 172.0 160.0
2.85
823
150.7 178.0 164.3
2.87
TABLE VI. Deuterium. I+' is the current to focus 4-/3, and /_' is the current to focus —/». IA— (I+-\-I~)/2. Currents in amperes.
B Field Current /"
349 400 440
A Field Current
/+' /-'
IA
66.3 73.2 69.7 68.2 78.8 73.5 72.4 82.4 77.4
0.854 .841 .848
results of similar measurements with deuterium. The final results do not have a precision as
high as would appear from these tables. First because an error in t h e determination of the peak current is magnified in the calculation of t h e moment. Second because of t h e possibility t h a t the check between t h e value of R as measured b y the ratio of currents and as calculated from the geometry m a y be partly the result of fortuitous compensation of errors in t h e very difficult geometrical measurements.
These measurements yield for the proton m o m e n t a value of 2.85 ± 0 . 1 5 nuclear m a g n e t o n s ; and for t h e deuteron a value of 0 . 8 5 ± 0 . 0 3 nuclear magnetons. Our j u d g m e n t of t h e precision comes from a discussion of t h e possible errors in the determination of the geometry.
DISCUSSION
The deuteron value should be somewhat more precise t h a n t h a t of the proton since the geometry
P H O T O E L E C T R I C P R O P E R T I E S OF ZINC
481
does not enter as importantly. This circumstance arises from the fact that the hfs separation for deuterium is so small that the moment in the B field is practically independent of t h e field. T h e ratio of t h e proton m o m e n t t o t h e deuteron moment is 3.35. This value should be somewhat more accurate t h a n t h a t of those for t h e individual moments since systematic errors should effect both values in the same direction. I t is to be noted that this value differs considerably from the value 4 obtained by Farkas and Farkas11 from t h e rates of t h e para-ortho conversion for hydrogen and deuterium.
T h e value of t h e proton moment which we previously obtained b y the use of atomic beams, 3.25db 10 percent is considerably higher than our
11 Farkas and Farkas, Proc. Roy. Soc. A152, 152 (1935).
present results. T h e cause of this discrepancy is rather obscure but may lie in our previous assumption t h a t the temperature of the beam was the same as t h e temperature of the source slit.
W i t h t h e sign of t h e moments established one can deduce an approximate value of t h e neutron moment on the naive assumption that the deuteron moment is t h e algebraic sum of t h e proton and neutron moments with the additional assumption t h a t t h e spin of t h e neutron is \. The neutron moment is thus — 2 nuclear magnetons.
In conclusion we wish to express our appreciation of t h e aid of a grant from t h e Carnegie Institution of Washington. Also, we wish to thank Professor H. C. Urey for the generous gift of t h e heavy water used in these experiments.
S E P T E M B E R 1, 1 9 3 6
PHYSICAL REVIEW
V O L U M E 50
The Photoelectric Properties of Zinc
CHARLES F . D E V O E , University of Wisconsin
(Received June 18, 1936)
The work function of a vacuum distilled surface of zinc was found t o be 4.24 volts at 1.5 X 1 0 - 8 mm pressure; as the pressure of air was increased the work function decreased to a minimum at about 10-6 mm pressure and then increased. Helium had no effect to a pressure of 3 mm. Nitrogen had no effect to a pressure of 10~3 mm.
EX P E R I M E N T S on the photoelectric properties of high melting point metals have shown that the metals must be heated at high temperatures for hundreds of hours before reproducible values for the long wave limits are obtained. Since this technique cannot be applied to zinc, several methods have been used to produce a gasfree surface rather than attempt to outgas a contaminated one.1-9 T h e values for the long wave limit of zinc obtained in these experiments ranged between 2940A and 4000A.
1 Richardson and Compton, Phil. Mag. 24, 575 (1912). 2 Hennings, Phys. Rev. 4, 228 (1914). 3 Kustner, Ann. d. Physik 46, 893 (1915). 4 Hennings and Kadesch, Phys. Rev. 8, 209 (1916). 5 Welch, Phys. Rev. 32, 657 (1928). 6 Werner, Zeits. f. Physik 67, 207 (1928). 7 Hughes, Phil. Trans. Rov. Soc. London 212, 205 (1912). 8 Dillon, Phys. Rev. 38, 408 (1931). 9 Rentschler, Henry and Smith, Rev. Sci. Inst. 3, 794 (1932).
Hughes7 measured t h e long wave limit of a v a c u u m distilled surface of zinc b u t t h e v a c u u m he used was low compared to that available now. T h e purpose of this experiment was t o s t u d y zinc surfaces prepared by Hughes' method under the best possible vacuum conditions.
APPARATUS
T h e experimental tube is shown in Fig. 1. T h e thin molybdenum strip A, spotwelded to a tungsten wire yoke and shaft, could be moved into t h e molybdenum collecting cylinder G, or in front of a tungsten filament E from which t h e strip could be heated by electron bombardment, or could be placed horizontally over either of t h e two quartz crucibles Z containing the zinc by means of an electromagnet acting on the soft iron b a r C. T h e collecting cylinder G was sup-