PHYSICAL REVIEW V O L U M E 140, N U M B E R 3B 8 N O V E M B E R 1965 Effect of Gravity on Gamma Radiation* R. V. POUND AND J. L. SNIDER Laboratory of Physics, Harvard University, Cambridge, Massachusetts (Received 26 May 1965) Recoil-free resonant absorption of the 14.4-keV y ray in Fe67 has been employed to measure the effect of gravity over a 75-ft vertical path in the Jefferson Laboratory, in an improved version of the experiment of Pound and Rebka. A Co67 source, initially 1.25 Ci, large-windowed proportional counters, and an enriched absorber foil 15 in. in diameter permitted a much increased counting rate. The employment of temperatureregulated ovens for source and absorbers and a redesigned monitor system to detect variations in waveform of the source velocity effected a reduction in systematic uncertainties. The result found was (0.9990±0.0076) times the value 4.905 X10~15 of 2gh/c% predicted from the principle of equivalence. The range given here is the statistical standard deviation set by the number of counts involved. An estimated limit of systematic error is 0.010. INTRODUCTION M ORE than fifty years have elapsed since Einstein proposed what has come to be called the "principle of equivalence," a generalization of the results of the experiments of Eotvos on the proportionality of mass and weight.1 Einstein proposed that no local experiment of any kind could distinguish between the effects of a gravitational field, on the one hand, and the effects of a uniform acceleration of the laboratory with respect to an inertial frame, on the other. In particular, if a source of radiation were to be viewed from a distance h below it, in a region having a uniform gravitational field of such strength that the acceleration of bodies in free fall were g, the observer should find the same properties that would be found if the whole laboratory were free of gravitation but were accelerating upward at the rate g. In this latter situation the velocity change taking place during the time of transmission of a given part of the radiation puts the observer at an effective velocity AF=gh/c. Therefore, one predicts that if a source of radiation at a height h above an observer were given an upward velocity gh/c, the effect of gravity would be cancelled. It is advantageous experimentally to seek the condition that would make an observer below the source see exactly the same properties he finds when making an observation from the same height above it. By such means, other distance-dependent factors in the experimental situation are removed. It is then the prediction that a source velocity difference equal to 2gh/c between the two arrangements should make observations in the two situations identical. In the above description no mention has been made of frequency or wavelength. The similar effect so intensively sought in stellar and solar spectra over the years is most often described in terms of the expected shift in the position of spectral lines, in the absence of a compensating velocity. So long as the velocity difference is small compared to c, its effect would be to produce, as * A short description of this experiment and its results has been published in Phys. Rev. Letters 13, 539 (1964). *A. Einstein. Jahrb. Radioakt. u. Elektronik 4, 411 (1907); Ann. Physik 35, 898 (1911). a Doppler effect, a frequency or wavelength shift amounting fractionally to gh/c2 for a one-way path. In a nonuniform gravitational field appropriate to astrophysics the argument is easily generalized to predict a fractional shift Av/vQ=—A=[(l-fF)/3^o]1/2X8Fi//3^, (8) where A7o is the total number of counts t h a t would have been collected in the absence of the resonant absorption. In terms of the time of the run r and the source strength R decays per unit time, this number is No=RTbtieD/4*. (9) From Eq. (8), for a given useful source strength and for F much smaller than one, the figure of merit for an example of recoil-free resonance is VH/F. The ease of achieving a large value of F with the Fe57 resonance line is the reason that it is as yet unsurpassed as a medium to detect small fractional shifts of frequency. Especially convenient is the fact that no low-temperature environment is required to obtain the high resolution. The precision of a given determination of QD is, beyond the statistical uncertainty, directly proportional to the precision of the quantity Vj to the extent that the conditions of Eq. (5) apply. In the interest of minimizing the statistical contribution to O,D from S, Vj should be large. On the other hand, too large a value of Vj complicates the procedure somewhat when high absolute precision is desired. One may carry out the analysis, including higher terms in the Taylor expansion, and obviously the result is dependent on the functional shape assumed for the resonance line. For lines of even symmetry only odd-powered terms contribute. For the Lorentzian of Eq. (1) one finds, still using VM2== VH*/3, oJ>'**aDll+i(yj/VH)*'], (10) where CLD is the result of applying Eq. (5), i.e., — (A+ + A " ) 5 , and aDr is the true asymmetry VD/C. An exact line profile would be needed to determine a practical correction factor to modify the linear example. A line profile determined in the early experiments,5 even though broadened considerably and measured with poor velocity stability, fit a Lorentzian curve surprisingly closely. A complication arises with the employment of Fe57 owing to the ferromagnetism of iron. The presence of an internal magnetic field and its resultant hyperfine struc- ture reduces the intensity of resonant absorption rela- tive to nonresonant absorption and therefore the mag- nitude of F that can be achieved without serious reduction in the counting rate. In addition the well- known six component structure of the radiation and correspondingly of the absorption can give rise to sig- nificant broadening of the central main resonance even though each component of the source radiation in principle resonates with its mate in the absorber, at E F F E C T OF G R A V I T Y ON G A M M A R A D I A T I O N B791 zero relative velocity. Such broadening can result from was recently reported by Cranshaw and Schiffer17 who small magnetic difference between pieces of the absorber performed an experiment at Harwell similar in most and of the source, slightly degrading the overlap of the respects but not including a monitor channel. components at a single velocity. The source strength desired could involve introduction into the iron of sufficient cobalt in itself to perturb the field in the source away from that of pure iron. Such internal field variation has been studied in iron-cobalt alloys.13 To avoid such problems it might be supposed that a nonmagnetic carrier would prove advantageous. However, although the lines in nonmagnetic stainless steels prove to be unsplit, they are so much broadened, probably as the result of irregularity in the local structures with resultant variations of isomer shift and/or quadrupole interactions, that in practice less resolution seems to be available than with iron. Unfortunately, the use of copper or platinum as a source matrix, although it results in a single relatively unbroadened emission line, does not lend itself to this application because a cor- THE NEW APPARATUS It was felt that a number of improvements could be made to enable a new experiment to be carried to at least an order of magnitude greater precision. Three main areas of reduction in systematic uncertainties were anticipated. First and foremost was the control and measurement of the important temperatures. Second, there was much room for improvement in the reliability and functioning of the "monitor" system that provided a way to eliminate from the result any perturbing effects in the source and its motion by obtaining data with an absorber close to the source. Third, basic improvements in reliability of the gating and switching functions of the electronics were possible. responding resonant, single-line absorber is not readily A result of some interest to the general problem of available. The best over-all combination appears still systematic errors in the experiment was reported in the to be an iron source and an iron absorber enriched in interim.18 This was the measurement of the rate of Fe57. An improvement by about two in the ratio of change of 7-ray frequency with hydrostatic pressure. resonant to nonresonant absorption is obtained through If sufficiently large, such an effect could necessitate the application of sufficiently strong magnetic fields to compensation or correction for the effect of the pressure polarize the source and absorber parallel to one another gradient in the atmosphere. The fractional change of and transverse to the direction of the radiation.14 Per- frequency was found to be about — 2.6X 10~18 per atmos- haps more important for this experiment is the fact phere. The atmospheric gradient should contribute less that permanent magnetsfixedto the source and absorber than 10~~5 as large an effect as gravity and is, therefore, units also reduce any possible effects on the resonance entirely negligible at the present level of precision. that could result from translation and reorientation of An important decision had to be made between the the system relative to the magnetic field of the earth. continued use of the short vertical height available in The experiment to be described here represents an extension to a higher precision of the earlier work of Pound and Rebka. That earlier work demonstrated conclusively the existence of the gravitational effect. In the course of that work the effect of temperature on the energy of the 7-ray emission and absorption lines was elucidated.15 With monitoring of temperatures to provide corrections, a measurement was made of the gravitational effect which agreed with the prediction to well within the uncertainties. The full data collected from that experiment had a statistical standard deviation of about 4% and an estimated limit of systematic error of about 10%.16 A result of 0.859 times the expected effect with an estimated over-all standard deviation of ±0.085 the Jefferson Laboratory or some other site. A vertical mine shaft could have advantages in thermal and vibrational stability and any increase in vertical path would reduce the importance of sources of systematic error accordingly. Even a taller building of stability comparable to our laboratory would therefore have advantages. An increase in vertical path would, however, with a given source strength, somewhat increase the statistical uncertainty because of the increased absorption in the path. The primary effect of the linear increase in the shift would be cancelled by the inverse square loss in intensity for a given absorber area and any further distance dependent decrease in intensity would represent a degradation. The contribution to the over-all statistical uncertainty because of the finite counting rate 13 C. E. Johnson, M. S. Ridout, and T. E. Cranshaw, Phys. Rev. Letters 6, 450 (1961); Proc. Phys. Soc. (London) 81, 1079 (1963). 14 G. J. Perlow, S. S. Hanna, M. Hamermesh, C. Littlejohn, D. H. Vincent, R. S. Preston, and J. Heberle, Phys. Rev. Letters 4, 74 (1960). 15 R. V. Pound and G. A. Rebka, Jr., Phys. Rev. Letters 4, 274 (1960). 16 R. V. Pound and G. A. Rebka, Jr., Phys. Rev. Letters 4, 337 (1960); R. V. Pound, Usp. Fiz. Nauk 72, 673 (1960) [English in the monitor channel would be reduced relatively. The logistical advantages of the Jefferson site proved so dominant that we decided to try to reduce the systematic uncertainties in that site, at least as a first step. That site provided a path of length approximately 25 m from the floor of the reverberation chamber in the basement to the ceiling of the penthouse above the main transl.: Soviet Phys.—Usp. 3, 875 (1961)]; The Mossbauer Effect, edited by D. M. J. Compton and A. H. Schoen (John WOey & Sons, Inc., New York, 1962), pp. 217-219; G. A. Rebka, Jr., thesis, Harvard University, 1961 (unpublished). 17 T. E. Cranshaw and J. P. Schiffer, Proc. Phys. Soc. (London) 84, 245 (1964). 18 R. V. Pound, G. B. Benedek, and R. Drever, Phys. Rev. Letters 7, 405 (1961). B792 R. V. P O U N D AND J. L. S N I D E R roof. Although the walls of this enclosed tower were originally entirely independent of the structure of the laboratory and therefore should, in principle, provide an ideal degree of isolation from vibration, the separation had been bridged in a few places in the course of remodeling operations over the years. The compressor unit of an A.D.L. Collins Helium Cryostat produced severe vibrations of the tower structure, apparently coupled through its foundations. The problem of regulating and monitoring the temperatures in the system was made especially severe through the absence of room heating or thermal insulation in the penthouse at the top of the system. An air-conditioning unit was installed there to try to limit the daytime temperature extremes in the summer months, but it proved to introduce such severe vibration as to render operation unreliable. An architectural drawing of the structure as it was employed is given in Fig. 2. To enable the precision of the result to be limited as nearly as possible only by systematic uncertainties and, particularly, to make possible tests with reasonably small statistical uncertainties, it was very desirable to improve on the operation obtained by Pound and FIG. 2. An architectural sketch of the arrangement of the system in the enclosed "isolated" tower in the Jefferson Physical Laboratory. Rebka. A direct step was an increase in initial source strength, by a factor of 5, to 1.25 Ci. Measurement standards in use at Nuclear Science and Engineering Corporation at the time led to the earlier source being mistakenly described as 0.4 Ci but it was closer to 0.25 Ci. As in the earlier experiment, the aperture available for the 7-ray beam was denned by a Mylar cylinder of 16-in. diameter filled at NTP with slowly flowing helium gas to avoid attenuation in air. An improvement in detector area over that of the seven scintillators, each 2f-in. in diameter, used by Pound and Rebka, was provided by a 90%-argon-10%-methane-filled proportional counter with a window 16 in. in diameter. After some unrewarding efforts to obtain improved net slope, F/VH, from various stainless steels, effort was concentrated on obtaining a large magnetically soft iron absorber foil enriched in Fe57. Over all these several improvements were expected to lead to a reduction by a factor of about 80 in the time required to obtain a given statistical uncertainty as compared to the original experiment. The radiation length for the 14.4 keV y ray corresponds to about 15 mg/cm2 of iron. For good efficiency a source should be thin compared to that or at least the activity restricted to a volume nearer than that to the surface. As a precaution against the perturbation of the internal field in so thin a source by too large a concentration of impurities, a large source diameter of 4 in. was chosen. The 1.25 Ci of cobalt was electroplated onto a foil of this diameter and of thickness 5 mg/cm2 made from iron enriched in Fe56 to 99.7% to reduce the resonant absorption by Fe57. The activity was diffused throughout the foil by annealing for 24 h at 1050°C. This source, prepared by Nuclear Science and Engineering Corporation, was received in March, 1963. About eight months of operations employed the source with an over-all system much like that of Pound and Rebka. Temperature-regulating ovens were used on the source, absorber and monitor units, and the improved aperture was obtained from the proportional counter mentioned above. The anticipated data rate was essentially achieved. Unfortunately, the internal consistency of data in a typical overnight run, from day to day or week to week, was not quite within expectations based on statistics. Almost all the data, when reduced to determine the effect relative to the monitor data correlating with inversion of the system, fell about 10% short of the predicted value. Had the internal consistency been up to expected standards, enough data were collected to yield a statistical standard deviation less than 0.5%. Considerable time and energy were expended trying to discover a spurious source of systematic error. There seemed to be evidence that the monitor channel was not providing information correctly related to that of the main absorber channel. A redesign of the basic elements of the system was undertaken in an effort to correct this situation. Several changes and improvements were made at this time, but the main source of error is believed to have been in the combination of an E F F E C T OF G R A V I T Y ON GAMMA R A D I A T I O N B 793 inadequately designed source mounting on the transducer with the sensitivity of the monitor absorber channel to source velocities in a direction transverse to the main beam, owing to its necessarily slanting angle period of the sine wave applied to the ferroelectric transducer centered about the extremes of velocity. A trigger pulse was generated from each direction of inversion of a monostable circuit driven from the audio of view of the source. The source, which had a Lucite oscillator through a network providing an adjustable backing disk, was found to be inadequately stiff and, phase shift. These pulses triggered a General Radio when driven at 117 cps from behind with a ferroelectric (GR) Model 1391-B pulse generator to form a pulse of cylindrical transducer of only 1J-in. diameter, there was about 3.S msec duration. An oscilloscope so set as to considerable excitation of modes other than the desired start a sweep on every third trigger pulse allowed accu- pure pistonlike motion. The redesign of the system in- rate equalization of the time intervals between succes- cluded both a very great reduction in any such spurious sive trigger pulses. A delayed and expanded sweep motions as well as a great reduction in the sensitivity allowed viewing of the pulse following that triggering of the monitor channel to transverse velocities. In addition, by enclosing the main absorber in a vacuum chamber, airborne sound was prevented from acting directly on the absorber foil. Integration of the monitor absorbers and counters mechanically with the source in a single rigid unit further reduced any vagaries originating in a difference in the methods and relative positions of mounting of these units at the two ends of the system. This new system was brought into operation in March, 1964, by which time the source had decayed to about 0.6 Ci. Operation with it was very much more consistent than before the modifications. It could be argued that the achievement of a very high data rate made possible the elimination of unanticipated spurious effects which might not otherwise have been found. the sweep. Because alternate sweeps were triggered by pulses derived from oppositely directed inversion of the monostable circuit, unequal time intervals resulted in resolved pulses when observed in this manner. Adjustment of the time constant for recovery of the monostable circuit allowed superposition of the triggers and thus equalization of the time intervals. The sine wave from the Hewlett-Packard 200CD oscillator operated at 73 cps was applied to the transducer through a step-up transformer. The phase of the gate pulses was set with the aid of another oscilloscope display. The vertical deflection of this oscilloscope corresponded to the sine wave and the horizontal deflection to the pulse from the GR pulse generator. That pulse was set to occur about 1 msec after the trigger pulse by use of the delay circuits in the pulse generator. This THE FINAL OVER-ALL SYSTEM allowed the square wave produced by the monostable A block diagram for the over-all system as finally evolved is shown in Fig. 3. Basically it is very similar to that of Pound and Rebka. Electronic gates accepted pulses occurring in each of the monitor and main channels in two periods, each in duration about 25% of the circuit to be used also as the drive to the electronic switches, since the reversals occurred outside the gate pulses that activated the counting circuits. Then the phase shifter was so set that the pulse occurring during the upward sweep of the oscilloscope was exactly super- HVDRAUUC MASTER -mzME—my={ SYNCHRO MOTOR FIG. 3. A block diagram of the over-all system. NOTE: PARALLEL SATURATING MAGNETIC FIELDS EXIST IN THE PLANES Of THE SOURCE AND ALL A8SQRBER FOILS. MONITOR CHANNEL B794 R. V. P O U N D AND J. L . S N I D E R posed on that during the downward sweep. The accuracy of the overlap was also a reasonable measure of the mean symmetry of the waveform of the sine wave during the gate pulses. The gating circuits themselves, for the main and monitor channels, were identical. The pulses from the single-channel analyzer in each channel were fed into a coincidence circuit together with the pulse from the GR pulser. The output of the coincidence circuit was fed through a Schmitt discriminator and the output of this in turn was applied to two coincidence circuits. Each of these was also supplied with a square wave synchronous with the sinusoidal modulation, one in each of the two polarities. These coincidence circuits provided the electronic switching. Thus the gated pulses from each, when phased properly, corresponded to oppositely directed velocities at the source transducer. The possibility exists that a small difference in dead time in the counters driven by the two sides of this circuit could lead to a spurious unbalance of the counts recorded on the two sides. To avoid this, pains were taken to make the dead time in the discriminator circuit preceding the switch longer than any later dead time. This was achieved by so connecting diodes to the capacitively coupled grid circuit that each pulse was followed by a long negative overshoot. A second pulse occurring before this bias-level overshoot had decayed could not reach the discriminator level. In this way a minimum time of about 3 /xsec was obtained between pulses and with scalers capable of 1-Mc/sec counting rates, no pulses should have been missed in the counters. This aspect of the electronics was subjected to repeated tests. Using regular pulses from a pulse generator, at rates up to 50 kc/sec, it was possible to examine the equality of the live time in the two counters corresponding to the two sides of a given channel. It was found that, with the symmetry of the trigger pulses set in accord with the criterion of superposition on the oscilloscope display, an inequality of about one part in 105 existed in the two times. The inequalities in the main and monitor channels were not identical, but changes in the trigger settings did change them by the same amounts. The inequalities were of a scale less than 0.25% of the asymmetry change expected due to gravity, on inversion. The stability and reproducibility of these inequalities were adequate to eliminate any significant contribution to systematic errors. That the dead-time situation was satisfactory was determined by operating both scalers of one channel from one side of the electronic switch, using random pulses at a very high counting rate near 5X104 per second which could be produced in the monitor channel with the pulse height analyzer running in the integral setting. The two scalers yielded numbers agreeing to within one count in 107 or more. Tests of the performance of the switching between sides with random pulse timing could only be carried out to limits set by statistics since the two sides counted independent samples. Little information on the performance of the system could be obtained in moderate times from such operation so the results with regularly spaced pulses were more informative on that point. Automatic alternation between oppositely directed values of the calibration velocity Vj was provided. The gate pulses were counted by a preset scaler. When the preset number, usually 44 002, was reached, the scaler interrupted the trigger pulses to the gate pulse generator and stopped the motor producing Vj. At the same time a pulse actuated a ratchet relay so as to reverse the motor that provided Vj and also to change over the output data to another set of four glow-transfer counters, which recorded counts after two decades of scaling ahead of the switches. After a 5-sec delay the motor was started. A time delay of about 25 sec allowed the backlash to be removed from the mechanical and hydraulic system. At the end of this time, the preset scaler was automatically reset to zero and another 44 002 triggers counted, and the cycle continued in this manner. The use of an even integral number of gate pulses to determine the period for a given sign of Vj avoided breaking into a gate pulse and any inequality of the number of pulses contributing to the two sides of a given channel. To allow study of the data accumulated in a given run, typically overnight, in a piecemeal manner, a photographic record was made during all runs. A ratchet relay and several time-delay relays made it possible to photograph automatically the eight six-digit glowtransfer counters and the preset scaler after every three, six, or twelve full periods of the cycling of Vj> corresponding to about 0.5-, 1-, or 2-h intervals. This record was particularly useful when a gross failure occurred in the system. Sometimes it enabled the salvage of some of the data prior to a failure. The pulses fed into the gating system were produced in the single-channel pulse-height analyzers of a Hamner 302 linear amplifier on the main channel and of a Hamner 328 plus 361 preamplifier in the monitor channel. The pulse rate in the monitor channel was nominally four times greater than that in the main channel and was limited mainly by electronic considerations rather than by solid angle in the detectors. There were two identical detectors for the main channel, one mounted permanently at each end of the path. Each was a proportional counter in the form of a 17-in.X17-in.X36-in. copper box with a 3-mil tungsten wire stretched along the long axis through Stupakoff seals used as guard rings. A 16-in. diam side window was covered with 1-mil Al foil inside a sheet of Mylar for strength. A preamplifier and cathode follower circuit were attached to one end of the box along with a pressure gauge. The counters operated with a 90% Ar-10% CH4 mixture at several ounces above atmospheric pressure, with about +3600 V on the central wire. At intervals of a few months the counters were flushed with fresh gas and then closed off again, but in general this E F F E C T OF G R A V I T Y ON GAMMA R A D I A T I O N B795 procedure made little apparent change in the detected spectrum. The counters were not rigid enough to allow evacuation so gas purity could only be obtained by sufficient flushing. These counters were an improvement over the set of scintillation counters previously used since their efficiency for detecting the 123-keV line accompanying the 14-keV line was considerably smaller. The ratio of the integral counting rate above the baseline to that in the 5-to-9-V window typically used for the 14-keV 7 ray was only about l.S to 1 for these counters. The net background within the window due to the 123-keV 7 ray and impurities was about 30% of the net total counting rate with the absorbers in place. The resolution of the counters was not very much better than that of the scintillation counters formerly employed. Furthermore, the resolution degraded significantly at high counting rates, and the gain decreased with increased counting rate. The elimination of the dependence of the relative weights of various parts of the absorber on the relative gains of the several multipliers used in the earlier experiment was considered to represent an important operational improvement in addition to the improvement from fuller utilization of the solid angle available in the system. The design of an oven suitable for the main absorber and capable of maintaining a constant temperature distribution, preferably a uniform one, was complicated by the necessity of maintaining a low absorption of the 7-ray beam over the whole area. With the exception of beryllium, no metal of sufficient thickness to provide significant thermal conductivity could be used to complete the enclosure across the ends. Figure 4 is a crosssectional drawing of the unit finally evolved. The absorber was a mosaic of foils each about 2 in. square cemented to a Mylar sheet of thickness 0.007 in. to cover the area of a circle of diameter 15 in. This sheet in turn was cemented to one face of an aluminum honeycomb support (Hexcel) 1 in. thick with hexagonal cells of diameter f in. The honeycomb cells were made of aluminum 0.003 in. thick. The purpose of the Hexcel was to provide both stiffening and transverse thermal conductivity with a minimum obstruction to the 7-ray beam. The support structure was found in practice to attenuate the beam by about 5%. This structure carrying the absorber was mounted across the central plane of an aluminum pipe of f-in. wall thickness and 24 in. length. The center 8 in. length of the pipe was separated from the end sections by 0.007-in. Mylar windows held with C -ring seals between flanges. The air pressure in this central section was reduced to a few microns by a mechanical pump connected to a pumping line which was then closed off and disconnected. In this way response by the absorber to airborne sound was very greatly reduced. That this was so was confirmed by tests with a vibration pickup mounted on the absorber surface. On the other hand the FIG. 4. The evacuated, regulated oven for the main absorber. The heating coils in the end spaces actually were wound with thin strips. absence of air increased the thermal inhomogeneity across the foil. The air spaces inside the pipe above and below the evacuated region were broken into three sections by thin Mylar diaphragms. The end ones of these were aluminized. Coils around the outside of the pipe provided heat and two thermistors connected in series, one mounted inside each end section, served as a reference for a Sargent Thermonitor control unit which supplied the current to the heating coils. A magnetic field to polarize the iron, everywhere greater than 50 Oe in the plane of the absorber and reasonably uniformly directed over the sheet, was provided by two banks of ceramic permanent magnets, one bank cemented to each of opposite sides of the pipe. Some evidence was obtained in the course of experiments leading to this final design that vibrations induced when the ac heater currents were used could cause difficulties. For that reason the power controlled by the Sargent Thermonitor units was rectified and smoothed by a transistor regulator. Six \% precision thermistors (Fenwal Iso-Curve glass probe thermistors, nominally 2000 fi at 25 °C) were cemented to the absorber surface at various azimuths and radii to allow sampling of the temperature distribution. Leads from these were brought through a multi- B796 R. V. P O U N D A N D J. L. S N I D E R lead seal on one side of the pipe. Experiments with the distribution and the regulation against ambient changes when the absorber was held at about 45 °C evolved the final system of heating. The heating current was passed also through additional end coils of aluminum strip wound on Mylar rings about 4 in. in diameter held between the two outermost diaphragms at each end of the pipe. The resistance of these coils, which determined the power in them relative to the power supplied to the coils on the pipe, was selected by experiment to provide a reasonably flat temperature distribution over the absorber. During operation the temperature of the absorber, in the form of the unbalance of a Wheatstone bridge set to the desired nominal resistance, was continuously recorded. Cnly one of the measuring thermistors was used, usually that one about midway between the center and the edge of the foil. A Brown strip chart recorder with a Honeywell 2HLA-7 preamplifier gave a full scale sensitivity of about 2°C. The timeaveraged temperature for a given run could be estimated to within less than 5 mdeg. The same leads from the bridge circuit and recorder were used for all locations of the absorber unit. These leads, consisting of a shielded pair, ran the full height of the tower with shorting plugs providing continuity at all sites except that in use. Thus the same net lead resistance and background ac pickup were always present. A similar arrangement was used to record temperatures for the monitor absorbers and for the source. The source and the monitor elements were put into an integrated unit of which Fig. 5 is a cross-sectional drawing. The two parts were connected by a Lucite ring so that the temperatures could be made different by adjustment of the relative power in the separate heating coils, both supplied from a single Sargent Thermonitor unit with rectified output. The reference of the regulator was a thermistor mounted near the monitor heating coils. The 4-in.-diam source disk was cemented to a 4§-in.diam copper disk § in. thick, and this was provided with braces of |-in.-thick copper for stiffening and mechanically coupling the disk to the l|-in.-diam, 2§-in.~long and ^-in.-wall cylindrical barium titanate transducer. At 73 cps, about 30 V rms across the thickness of this transducer provided the velocity modulation to the regions of maximum slope in the absorption line. The transducer in turn was mounted on a heavy brass base which was fastened to the piston of a lf-in.-diam hy- draulic cylinder, to provide the calibrating motion. This was driven by a f-in.-diam master cylinder moved by a 1.5 rpm synchronous motor rotating a JX20 nut on a screw. A precision thermistor at the center of the source disk provided for temperature monitoring. The outside of the source oven was wrapped with a 1-mm thickness of lead to reduce the level of external radia- tion. Permanent magnets were cemented to the outside of the assembly to provide a transverse field in the plane of the source of about 100 Oe. The monitor part of the assembly was designed around a pair of Reuter-Stokes proportional counters. Each was provided with a monitor absorber foil of 43.5% Fe57 and 1.8 mg/cm2 iron cemented to a beryl- lium sheet about 0.030 in. thick. These absorbers were fastened over holes in a copper plate on opposite sides of the beam of radiation along a diameter perpendicular to the magnetic field at the source. A magnetic field parallel to that in the source was provided in the plane of the absorbers by magnets cemented to the copper mounting plate. The proportional counters were wrapped in lead and considerable lead screening was provided to limit the radiation into the proportional counters to that passing through about |-in.-diam circles of the absorbers. The actual monitor aperture was limited at the counter windows to give as closely as FIG. 5. The oven containing both the source and its ferroelectric transducer and hydraulic cylinder as well as the monitor absorbers and their associated proportional counters. A yoke and set of coil springs that kept the hydraulic cylinder under pressure have been omitted to avoid complication. possible uniform weight to all parts of the source in each counter. The first tubes tried were filled with a 90% krypton-10% methane mixture and became useless in about 24 h of running in the intense radiation near E F F E C T OF G R A V I T Y ON GAMMA R A D I A T I O N B797 the source. Later tubes were filled with 90% krypton10% nitrogen. An improvement in the lead shielding and reduction in the operating voltage from 2300 to 1900, made possible by increased preamplifier gain, may have been the main cause of the improved life. The pulse-height spectra from the two counters were made equal by adjustments of apertures and of supply voltage. The krypton-nitrogen-filled counters seemed to degrade quickly after about three months of operation. The gain fell off and the spectrum broadened. Two replacements were installed in the interest of maintaining symmetry in the course of the experiment. The monitor temperature was sampled by a set of four precision thermistors, two on each absorber, connected in series-parallel to give a nominal resistance equal to that of a single thermistor unit. The resistance of the combination was believed to represent well an appropriately weighted average. The monitor temperature proved to be very stable, hardly ever changing by more than 0.010 deg, once the regulator was set for a given run. At the high counting rates in the proportional counters of the monitor, the dependence of the counter gain on intensity gave rise to an anomalous dependence of sensitivity on setting of the window of the pulseheight analyzer. The variations of counting rate accompanying the sinusoidal modulation correlated with a variation of position of the pulse height spectrum. Selection of only the low-half of the spectrum resulted in almost the full available change of counting rate for a given change of source velocity. The statistical standard deviation was only that of half the counts. In effect, the counts in the upper-half of the pulse-height distribution were unaffected by velocity and conveyed no information on source velocity. This phenomenon was discovered when short runs were made with different pulse-height-analyzer settings in search of optimum sensitivity. The behavior was as if there were radiation on the high side of the spectrum not participating in the resonance, but the correct explanation was confirmed by observing the pulse-height distributions in coincidence with a short gate pulse, as a function of the position of the gate pulse in the velocity-wave form. The result was that the changes in intensity accompanying the varying absorption produced spectra having identical counting rates on the high-voltage sides but differing on the low sides, as sketched in Fig. 6. Although improved statistics were available through this effect, it was not used for fear that the exact information sought from the monitor channel could be degraded. In particular, if the gain of the proportional counters had a lag not negligible compared to the duration of the gate pulse, as might be expected if the effect results from a charge plasma sustained in the counter, a distortion in the effective gating function would be introduced. To avoid complications of such origin, the window in the monitor channel was set to accept the full pulse-height peak. COUNTS 14.4 KEV PEAK — MOVING SOURCE - - ZERO SOURCE VELOCITY CHANNEL NUMBER - * - FIG. 6. Curves representing pulse-height distributions from the proportional counters in the monitor channel at two source velocities. The lower peak counting rate at the null point in the velocity waveform of the source correlates with a larger gain in the counter. The two curves overlap on the high-voltage side and thus an artificially large variation of counting rate with source velocity results in the lower voltage channels. SYSTEM OPERATION The data taken with the apparatus in its fully evolved form was composed typically of overnight runs and full time runs over weekends. Severe vibrations induced by the nearby helium liquefier made weekday runs untrustworthy. For each run 11 numbers were obtained. Four of these corresponded to counts at the four combinations of the source velocities dtVM and zLVj for the main channel and four more for the monitor channel. The other three numbers corresponded to the averaged deflections of the recorded traces for the measuring thermistor bridges connected to the source, the monitor absorber and the main absorber, respectively. Following the reasoning above, the four counts for a given channel were used to calculate, from the measured value for Vj, the effective asymmetry of the absorption line, with its statistical standard deviation, for the main channel and for the monitor channel. Each of these was then corrected to a fixed thermistor resistance of both the source and the respective absorber. One complication involved the fact that the monitor observed the source at a slant angle. Doppler effects due to motion along the main axis of the system were reduced by the cosine of the slant angle to the monitor. Thermal shifts were, of course, absolute and independent of the angle of view. In an earlier form no provision for accurate recording of the source temperature was provided on the grounds that changes of source temperature affected the main absorber and the monitor asymmetries equally and so cancelled from their difference. On the other hand variations of transducer wave form would affect only the velocity along the axis of the system and their effects at the monitor absorber would be reduced by the same slant projection as was the effect of the calibration velocity Vj. Since a crucial function of the monitor was to correct for variations of transducer wave form, particularly any that might correlate with inversion, recording of source temperature was added and correction of asymmetries to a standard source temperature was B798 POUND AND J. L. SNIDER FIG. 7. The difference between the temperature-corrected asymmetry of the main channel . In practice the modulation amplitude was chosen by experimentally minimizing S, and at that setting the transmission averaged over the quarter cycle of sine wave was increased about 10% above that at no modulation. For a Lorentzian line, the coefficient of the correction term in an expression like that in Eq. (10) but carrying VM as a parameter drops to zero at VM*= VH2, which also corresponds to operation at the half-value point in the absorption curve. The point VJI^—IVH2 corresponds to operation at f of maximum absorption in a Lorentzian line, but the line was less than 40% deep. It is thus indicated that our usual operating point, averaged over the actual waveform during the period of the gate, is between the point VM2= VH2/^ and VM2== VH2, and thus an averaged correction factor would fall between 1+f (VJ/VH)2 and 1. For our values F/=7.83X10"4 cm/sec and Vj~1.5 X10-2 cm/sec, (Vj/VH)2~2.7XlO-\ Accordingly, it seems reasonable to make as a correction to all values an increase by 0.002 and to include a systematic uncertainty of ±0.002 to cover the range between no correction and the full correction that would apply if operation had been at the point of maximum slope on a Lorentzian line. The correction for nonlinearity and a systematic uncertainty relating to it were not included in the original brief publication of these results.19 The height of the path was measured several times with a steel tape and is believed known to one in 2000. Uncertainties in g and c, using standard tabulated values, are negligible. C ne is led then to the final result, including the fulllength and half-length runs weighted according to the inverse square of their statistical standard deviations, AV/c= (0.9990±0.0076)2 gh/c,2 19 R. V. Pound and J. L. Snider, Phys. Rev. Letters 13, 539 (1964). B802 R. V. P O U N D AND J. L. S N I D E R TABLE II. A list of estimated limits of contributions to systematic error. Origin Path length Temperature coefficient Temperature distribution Velocity calibration Nonlinearity Simple sum Amount ±0.0005 ±0.0005 ±0.0050 ±0.0020 ±0.0020 ±0.0100 where the range indicated is the over-all statistical standard deviation but does not include the systematic uncertainties. In addition, the central value may range over ±0.010 as our estimated limit to the systematic errors, all simply added arithmetically. The several contributions to this are listed in Table II. If one were to follow the frequent practice of treating such uncertainties as independent and of random sign, an rms value of ±0.006 would result, and taking the rms value of this and the statistical standard deviation yields an overall uncertainty of ±0.0095. It seems fair to describe the present result as confirming the prediction based on the principle of equivalence to a precision of about 1%. The approximately 1% limit of error in this over-all result represents a fractional frequency change of 5X10~17, or a claimed precision of 5X10"5 times the basic resonance linewidth. Such a fine-scaled " splitting'' of the line was made possible by the fact that inversion allowed separation of the effect sought from inherent shifts. Furthermore, the use of the monitor channel eliminated the effects of variations of transducer waveform. Such variations, without a monitor, play a role analogous to cavity pulling in a maser oscillator or background effects in a densitometer analysis of a spectral line. There might be some interest in a " one-way'' result, that is a measurement that would be sensitive to a shift dependent on height or distance but not reversing sign with inversion, which would then cancel out of the "two-way" measurement. Unfortunately, in a one-way experiment one has no independent test for the shifts that might accompany the necessary geometrical changes involved. The data of greatest reliability bearing on this question come from combining the results of the final apparatus, using the full length, with those for half the length of the path, with the source kept at a given end. For the experiments over half the path, the detector was moved with the absorber, so essentially the whole area of the absorber was in use. Even so the geometry was necessarily different for the two situations. For example, the Hexcel support, with the larger spreading angles in the shorter runs, shadows a different part of the absorber, than in the full-length run. A change in the effect of any inhomogeneities in the absorber is therefore introduced. The experiments were not carried out with such a comparison in mind, so there were long separations in time between the recording of some of the data involved. Nevertheless if one looks at the change in the net asymmetry difference between the main and monitor channels, when the absorber was moved from the middle of the tower to the top, keeping the source at the bottom, the result was about 0.87 times the 1.338X10"15 expected value of gh/c2, corresponding to the part of the tower added on a one-way basis. Similarly, when comparing data with the beam going downward, a result about 1.15 times the same expected value was found, corresponding to moving the source from the top to the center of the tower, with the absorber at the bottom. In each case statistical standard deviations were about ±0.06. These results indicate a frequency increase of about 2X 10~16 associated with this lengthening of the path. This may have resulted from the changed geometry. A change of about 0.1 °C in the weighted mean temperature would be necessary to produce such a shift, and that was quite a large part of the total range of temperatures indicated by the six thermistors. Some consideration is being given to experiments over longer baselines to check further on this point. With the completion of this phase of the experiment, one is led to wonder how it could be further improved. The role of thermal instability as the major source of systematic error could be very much reduced by provision of an environment itself stabilized thermally. Space in either a well-controlled building or an underground facility would go a long way toward eliminating this problem. Of course, a longer baseline would reduce the importance of a given absolute uncertainty in direct proportion to its total path. The uncertainties of velocity calibration and of nonlinearity would not be affected by a moderate increase in baseline. Hydraulic pistons of much better quality could undoubtedly be found or made. With a longer baseline, however, the net shift begins to compare to the maximum usable calibration speed. Consequently it is attractive to incorporate a third velocity into the system. One could use a platform, for example, at the top of the system, continuously moving upward at a precisely known speed near gh/c, to compensate for the gravitational effect continuously. Such a velocity corresponds to about 1% of the path h in 3 days, so that system would need to be recycled frequently to prevent important geometrical changes. The velocity measurement on which the result would depend most would then be that of this compensating motion. The cycling motion Vj would be used only to measure any discrepancy between the shift and the compensation chosen. Correspondingly, the uncertainty due to nonlinearity would also enter only in this less important way. To justify a renewed effort, one must, of course, reduce the statistical uncertainties accordingly. Using a E F F E C T OF G R A V I T Y ON GAMMA R A D I A T I O N B803 source strength renewed to our original 1.25 Ci and counting with our present absorber aperture, it would be difficult to obtain sufficient counts to reduce the statistical uncertainties below about 0.25%. Of course with a larger absorber one could gain accordingly in this area. The possibility, a priori, of a discrepancy between the simple theory and experiment in the domain between a few tenths and one percent does not appear large. One aspect that may be worthy of mention is that g includes the effect due to the earth's rotation amounting to a reduction from the purely gravitational acceleration by about 0.3%. Agreement at that level would then compound the effects of gravitation and of central acceleration. These latter have, of course, been observed directly, in their own right, in experiments with rotating systems.20,21 The view that the local time scale depends on gravitational potential appears to require a coherent source for confirmation. The present experiment is unable to distinguish between frequency changes and velocity changes, for example. It appears as if an experimental 20 H. J. Hay, J. P. Schiffer, T. E. Cranshaw, and P. A. Egelstaff, Phys. Rev. Letters 4, 165 (1960). 21 D. C. Champeney, G. R. Isaak, and A. M. Khan, Nature 198, 1186 (1963). comparison of clocks at different potentials would make a useful complementary contribution to the over-all status of confirmation of theory. ACKNOWLEDGMENTS Many persons have contributed to the design and carrying through of this experiment. Dr. Glen A. Rebka, Jr., who collaborated with one of us in the original version, contributed through that work in a most fundamental way. We are indebted to Dr. R. W. P. Drever who, during a one-year residence here, designed and introduced our large proportional counters and contributed in many less concrete but important ways. The members of the shops and services in these laboratories, and particularly F. B. Robie as Engineer in the Department of Physics, have aided generously in many indispensable ways. Paul Horowitz rendered helpful and efficient aid during a couple of summers. Finally, Rufus Walker, Jr., and Jack T. Sanderson, at the time graduate students working on other projects, gave generous and valuable help on several occasions, in particular by designing a computing program. Thanks are due to the U. S. Office of Naval Research for support through Contract No. Onr 1866(19) and to the Francis Barrett Daniels Memorial Fund which supported the purchase of some of the equipment. - — W m m m SLAVE CYLINDER LEAD COLLIHATING SHIELD FIG. 5. The oven containing both the source and its ferroelectric transducer and hydraulic cylinder as well as the monitor absorbers and their associated proportional counters. A yoke and set of coil springs that kept the hydraulic cylinder under pressure have been omitted to avoid complication.